EPA \ TAQMN \ Air Quality Monitoring \ TAQMN site \ Instruments

Instruments

Note: This function or related web page display of the equipment brand, model system for comparative reference data application, we have stringent quality assurance operating procedures in order to maintain data quality, but does not ensure that all the same instrumentation, model of the quality or goodwill , all equipment operators shall not be any use of this site's content for business activities.

name model specification
Sulfur Dioxide Analyzer Ecotech 9850
  1. Range 0~20000ppb
  2. Linearity error: 1% of full scale
  3. Lower detectable limit: 0.5ppb
  4. Precision: ±1% of reading
  5. Zero Drift: 30 days less than <1ppb
  6. Span Drift: 30 days less than 0.5% of reading
Nitrogen Oxides Analyzer Ecotech 9841
  1. Range 0~1000ppb
  2. Linearity error: 0 ppb
  3. Lower detectable limit 0.5ppb
  4. Precision: ±1% of reading
  5. Zero Drift: 30 days less than 1ppb
  6. Span Drift: 30 days less than 0.5% of reading
Ozone Analyzer Ecotech 9810
  1. Range 0~100ppb
  2. Linearity error 0 ppb
  3. Lower detectable limit 0.5ppb
  4. Precision: ±1ppb of reading
  5. Zero Drift: 30 days less than 1ppb
  6. Span Drift: 30 days less than 0.5% of reading
Carbon Monoxide Analyzer Horiba  
Gas Dilution Calibrator Model 4010
  1. Zero gas 0~20 L/min
  2. SCCM SPAN Gas 0~200
  3. Adjustment range 100:1
  4. Permeation drier stability ±0.1℃
  5. Ozone Generator 5ppm/L
Zero Gas Generator Sabio 1001  
Hydrocarbon Analyzer Dani 462
  1. Measuring Range 0~50ppm
  2. Linear: less than 1% of full scale
  3. Precision: 1% of full scale
  4. Zero Drift: 24hours ±0.2% of full scale
  5. Span Drift: 24 hours ±1% full scale
Suspended particulate analyzer Verwa 701  
Temperature/Relative Humidity Probe Metone 083C
  1. Temperature range -20℃~60℃
  2. Temperature error ±2%
  3. Measuring Range of Relative humidity 0~100%
  4. Relative humidity error ±2%
Anemometer Metone 014A
  1. Range 0~10 m/s
  2. Starting threshold 0.25 m/s
  3. Error: 1.5% of full scale
  4. Linear: 0.5%
  5. Operating Temperature range -50℃~70℃
Weather vane Metone 024A
  1. Measuring range 360°
  2. Starting threshold 1.0 m/s
  3. Error ±5%
  4. Linear 0.5%
  5. Operating Temperature range -50℃~70℃
Automatic Acid Rain Analyzer Ogasawara us-760S
  1. PH range 0~10
  2. PH error ±0.1PH
  3. EC range 0~1000 us/cm
  4. EH error ±5%
  5. Rain temperature 0~50℃
  6. Rain temperature error ±0.5℃
  7. Rainfall 15.7ml